-40%
Partlow MRC 5000 Temperature Recorder, 1 Ch
$ 800.44
- Description
- Size Guide
Description
Untitled DocumentPartlow MRC 5000 Temperature Recorder, 1 Ch
Model:
MRC 5000
| Order No:
51000011
Features 10' diameter recording charts with user selectable chart speeds: 8, 12, 24, and 48 hours, and 7 days. 1 trend pen version, with configurable spans.
Brand:
Partlow
Model:
MRC 5000
Order No:
51000011
Condition:
NEW
Description
Partlow MRC 5000 Temperature Recorder, 1 Ch
Features
Two programmable alarms per pen
Universal sensor inputs: thermocouple, 3-wire RTD, or DC linear (mA, mV, V)
Just one one jumper to configure each input port
4 digit LED display and 3 button keypad make configuration changes fast and easy
User interface provides access to test and calibration facilities with dual function MODE switch
Optional Configurator port for ultimate programming ease, using the optional PC program, parameters can be modified, saved, and downloaded in seconds
High/Low limit capability available
Universal power supply
NEMA 3 case standard
Fits 12.7" square panel cutout
Recording charts supplied in boxes of 100 are sold separately
Included
Partlow MRC 5000 Temperature Recorder, 1 Ch
Specifications
Recording
Pen Type
Disposable fiber tip
Pen Colour
Pen 1 - Red; Pen 2 - Green
Chart Size
10"
Chart Drive
Stepper motor
Chart Rotation
User configurable; 8-hour, 12-hour, 24-hour or 7-day
Chart Span
Bottom & top of span, -9999 to 9999 units
Input Ranges
J T/C
0 to 700°C & 0 to 1400°F
K T/C
0 to 1360°C & 0 to 2500°F
T T/C
-200 to 400°C & -330 to 750°F
R T/C
200 to 1650°C & 400 to 3000°F
S T/C
200 to 1650°C & 400 to 3000°F
RTD 100Ω
-140 to 400°C & -220 to 750°F
Platinum
0.00385Ω/Ω/°C
DC Current
0 to 20mA, 4 to 20mA; Internal 4.7Ω shunt resistor
DC Voltage
0 to 25mV, 0 to 50mV, 10 to 50mV, 0 to 5V, 1 to 5V
Input Performance
Measurement Error
Type J, K, T, R, S & RTD: ±0.25% of span ±1°
mA, mV & VDC: ±0.25% of scaled span + 1 LSD
Cold Junction
± 0.2°C @ 25°C
Compensation Error
0.04°C/°C deviation from 25°C
Linearization Error
TCs: ± 0.25°C typical. ± 0.5°C worst case
RTDs: ±0.1°C typical, ± 0.3°C worst case
Ambient Temperature Error
±0.01% of span/°C deviation from 25°C
Alarms
Number
Up to 2 process alarms for each of 2 inputs
Type
Process high or low
Limit Devices
Optional high/low limits for each input with latching output
Normally open output latches open
Red reset button included to right of display
Hysteresis
Fully adjustable, 0 to 200 units, single sided
Security
Alarm setpoint changes can be prohibited
Sensor Fault Action
Alarms work normally in "Hi" and "Lo" conditions
Relay Outputs
Relays
SPST, contacts rated 5A resistive at 115VAC, 2.5A resistive @ 230VAC, 1/8 HP @ 230VAC (single phase), 250VA A 115/230 VAC
Power Requirements
Line Voltage
90-264 VAC, 50/60 Hz
Power Consumption
18 VA max.
Construction
Enclosure
Injection molded Noryl case and cover with acrylic window
NEMA Rating
NEMA 3 standard
Mounting
Panel or wall
Panel Cutout
12.7" x 12.7" (322.6 x 322.6mm)
Panel Depth
2.5" (63.5mm)
Panel Protrusion
1.3" (33.0mm)
Overall Dimensions
14 x 14 x 3.8" (355.6 x 355.6 x 96.5mm)
Weight
15 lb (6.8 kg)
Terms & Conditions
Item Condition:
Unless specified otherwise all products listed are Brand New in the original factory packaging and include a full factory warranty.
Shipping:
We currently only ship to U.S. buyers. We ship items within one or two business days from payment. We do not process shipments on weekends and holidays. Local pick up is not available. We only ship to addresses verifiable by either PayPal or credit card issuing bank.
Please note that we cannot alter the address once a purchase is made. In the case of PayPal payment, we will ship orders when the payment is marked “Eligible” for PayPal Seller’s protection.
*Shipping Rates are for Continental US, Please Contact us for other US Destinations
Payment:
Currently all payments must be made directly though PayPal
Sales Tax:
We are required to charge sales tax in North Carolina
Return Policy:
If for any reason you are not 100% satisfied with a product you receive from us, you may return the regular stock product within 30 days from the time of receipt, for a full refund or exchange of the product you purchased (less any shipping costs). Within this time, products may be returned; freight prepaid and must be in their original packaging and in resalable condition.
Shipping errors, missing items or damages must be reported to us immediately (damages should be accompanied by a signed bill of lading acknowledging the damage). Items returned will be inspected for damage or misuse at which time they may be subject to restocking fees, other fees or refused, prior to the issuance of any credit.